Condicionador de sinais para termopar tipo K

From thermocouple millivolts to a 10 mV/°C output, with cold-junction compensation

  • Electronics
Year
2026
Context
Signal Processing course — Fundação Liberato
Team
Maurício Gedrat, Leonardo Kohn

Illustrative background photo: Anne Nygård / Unsplash

Overview

A circuit that turns a type K thermocouple's voltage, from −10 to 120 °C, into a linear 10 mV/°C output.

The problem

Across −10 to 120 °C the thermocouple changes by only 5.312 mV. It also measures a temperature difference: without knowing the cold-junction temperature, the reading is meaningless.

The approach

  1. 01Output scale chosen first: 10 mV/°C, 0 to 1.300 V, easy to read on a multimeter or an A/D converter.
  2. 02INA128 instrumentation amplifier for the small differential signal, with a 249 Ω RG (gain ≈ 201.8).
  3. 03An LM35 measures the cold junction, and a TL071 adds the compensation and trims zero and gain.

How the system works

  1. 01

    Type K thermocouple

    5.312 mV over range

  2. 02

    INA128

    G ≈ 201.8

  3. 03

    LM35

    cold junction

  4. 04

    TL071

    zero and gain

  5. 05

    Output

    10 mV/°C

Implementation

  • Two-point calibration to trim zero and gain independently.
  • Uncertainty analysis combining error sources by root sum of squares (RSS).

Result

Estimated total uncertainty of 0.59%, under the 2% limit the brief required.

0.59%
total uncertainty
10 mV/°C
output scale
−10–120 °C
design range